| |
|
|
|
|
|
|
Labeling methods |
| |
|
Deuterium Introduction |
For more information see the
publication below as well as the
Publications page:
Engen, J.R. & Smith, D.L. (2000). Investigating
the higher order structure of proteins: Hydrogen exchange, proteolytic
fragmentation & mass spectrometry. In "Protein and Peptide Analysis: New
Mass Spectrometric Applications" (J. Chapmann, ed.). Meth. Mol. Biol.
Vol. 146.
|
|
Deuterium can be introduced to the
protein sample in a number of ways. Two primary methods are dilution
and gel filtration. It is always recommended that deuterium is
introduced into a protein sample that is already in solution. Addition of
D2O to lyophilized protein complicates the data at early
exchange-in times due to solvation effects.
|
|
Dilution |
|
With the protein in a buffer of
H2O, the sample is diluted by adding an excess of the same
buffer but made with D2O, in a dilution factor of 15-20 fold
by volume. This effectively raises the deuterium level to greater
than 95%. Dilution is an easy method but requires more protein
since the concentration is significantly reduced. |
| |
|
Gel
filtration |
This method works by introducing the
protein in a buffer of H2O into a small spin column that has
been equilibrated with the same buffer made with D2O.
The spin column is filled with G-25 gel filtration media. After a
brief spin in a table-top centrifuge, the protein elutes out the column
into a receiving tube. The protein is now in D2O and
the H2O is trapped in the gel filtration media. Using
this technique requires precise timing of how long it takes to elute the
protein from a specific bed volume at a specific centrifuge speed.
A simple test can be done with a visible protein like cytochrome c,
myoglobin, or green fluorescent protein to determine the optimal
parameters.
|
| |
| Pulsed
vs. Continuous-labeling |
Information below was summarized
from this publication:
Deng, Y, Zhang, Z. and Smith, D.L. (1999). Comparison of continuous and
pulsed labeling amide hydrogen exchange/mass spectrometry for studies of
protein dynamics. J. Am. Soc. Mass Spectrom. 10,
675-684.
|
|
|
The actual labeling step, when the
protein is in deuterium, can also be done in several ways. Two
primary methods are
pulsed-labeling or
continuous labeling (see Figure 1)
click for a
larger image

Figure 1
|
|
Pulsed-labeling |
|
the protein sits under specific
conditions for a period of time (i.e. in denaturant for 30
minutes). Then a short pulse of deuterium is introduced and the
reaction is quenched. |
| |
|
Continuous-labeling |
|
the protein is exposed to deuterium
at time=0 and left in deuterium. At a various times the reaction
is quenched by moving an aliquot of sample to quench
buffer. |
| |
The primary difference between the two
methods: In continuous labeling, a cumulative summary of the
populations of the molecules if evident in the spectra. In
pulsed-labeling, an instantaneous snapshot of the protein populations is
evident. Continuous labeling is useful for looking at populations of
protein molecules under conditions where the folded state is favored since
the minor contribution of the unfolded states is integrated over
time. Pulsed-labeling is useful for measuring the unfolding and
refolding rate constants in conditions where the unfolded state has been
artificially increased by addition of chemical denaturant, heat or pH
changes. |
| |
Two spectra are shown at
the left. They are from the same peptide of a larger protein
that was labeled by either (A) continuous or (B)
pulsed-labeling.
In (A), the protein hydrogens were available
for exchange with deuterium for 30
minutes before being analyzed. In (B), the protein was
equilibrated in 3M
urea for 30 minutes, was pulsed for 10 seconds with deuterium at pH
7 and then analyzed.
From the spectra in (A), the cumulative
population of molecules that have unfolded is evident, since once
the molecule unfolds, it becomes labeled (a process that cannot be
reversed since the concentration of deuterium is >95%). In
(B), there are 2 populations of molecules, indicated by the two
isotope distributions. The higher mass distribution represents
the percent of the population that is unfolded after 30 minutes in
3M urea. |

Figure 2 |
|
| |
| Quench-flow and rapid
mixing |
|
Rapid mixing systems, such as the
Biologic quench flow system or other such systems can be used to
investigate events that are as fast as 10 milliseconds. Things under
this category can include rapid exchange of surface amide hydrogens or
protein folding events.
Protein folding can be studied with a
type of pulsed-labeling. A denatured protein is allowed to refold by
dilution of the denaturant. After certain refolding times, the
protein is pulsed with deuterium and the reaction quenched. Analyses
of this type can be automated with quench-flow devices.
An example of a set-up that can be used
is shown in Figure 3:
click
for a
larger image

Figure 3 |
| |
|
|
Which ionization method to chose? |
| |
|
Currently, there
are two primary ionization methods for doing hydrogen exchange
analysis:
Electrospray (ESI)
MALDI
In the past, FAB was also
used.
Electrospray
- HPLC step washes away all
the labile hydrogens at side chain positions
- Temperature and pH are
easily controlled
- When coupled to HPLC,
large, complex digests can be analyzed
- Samples must be desalted
and free of impurities
- Multiple charge states can
be complicating
MALDI
- Potential for higher
throughput
- Deuterium losses can be
harder to control
- Generally easier to do
than ESI
- All data are aquired in a
single spectra
- Sample salts and
impurities are much more tolerated
- Lack of multiple charge
states make spectra simple
People using MALDI:
Komives Lab,
UCSD
Use of MALDI to look for
binding surfaces
(PNAS 95, 14705-14710)
Fitzgerald Lab, Duke
SUPREX technique to
analyze protein stability
(PNAS 97, 8296-8301)
|
| |
|
|
The
basic experiment |
| |
|
For a continuous labeling
experiment, a protein is labeled with
deuterium at pH=7.0. The reaction is quenched by lowering the
pH to 2.5 and the temperature to 0 deg. C. Its VERY important
that that the pH change is precisely controled. The best way to do
that is to use a weak (25 mM) phosphate buffer for the labeling step
at pH =7.0 and then use a strong (100 mM) phosphate buffer for the
quench step. This ensure that the pH will drop to 2.5 and stay
there. See below "Controls &
Calculations".
To do electrospray (see the
HX MS background page as well as
above) mass spectral analysis, the sample must be desalted before it
enters into the instrument. Desalting serves a second, and
very important purpose - it washes away the deuterium from amino
acid sidechains (see Figure 4). The exchange rate of the amide
hydrogens in side chains (for example in arginine, or histidine) is
very-very fast compared to the exchange rate of the amide hydrogens
in the backbone. When desalting is performed with standard
HPLC buffers (which have a pH of 2.5 and are made of H2O), the
hydrogen in the water of the HPLC buffers replaces the deuterium
that exchanged into the side chains. This leaves only
deuterium label at the backbone amide positions. Having only
these amide hydrogens to analyze is a convenient situation
1. it gives an
individual exchange-rate sensor for every amino acid (expect
proline).
2. it eliminates messy details of exchange
into side chains, which can be influenced by
things other than
secondary structure and solvent shielding.
Figure 5 shows an HPLC column,
injectors and associated tubing in a bath of ice water to maintain
the temperature at 0 deg. C. Using this set-up reduces the
losses of deuterium during analysis, whether it be from a whole
protein or from peptides.
Whole proteins can be analyzed, or
digested into smaller pieces with an enzyme. The enzyme
that seems to work the best is porcine pepsin. It likes
working at pH 2.5 and can tolerate being at 0 deg. C. When the
sample is ready for analysis, it is digested for 5 minutes at 0 deg.
C with pepsin or it can be digestion inline immediately prior to RP-HPLC
separation (see Figure 5). After digestion, the peptides are separated
quickly at 0 deg. C in 5-6 minutes. The column, solvent lines
and injector are kept in an ice bath (see Figure 5). Figure 1
panel B illustrates a brief flow-chart of a continuous labeling experiment.
|
click for a
larger image

Figure 4
click for a
larger image

Figure 5 |
|
| |
| Other
experiments |
| |
|
There are many variations one can do to
the general type of experiment explained above. Whether analysis if
done with MALDI rather than electrospray, whether a pH pulse is used,
whether denaturants are thrown in.... all experiment must take into
account the pH factor and the temperature factor (see
Background/Theory page). Without doing so,
the results of such analyses can be meaningless.
It is also critical to USE PROPER CONTROL
samples. The details of this are discussed in the section titled
"Controls and Calculations" below.
|
| |
|
|
Controls and
calculations |
| |
| Adjustment
for deuterium loss |
| |
Although some loss of deuterium label
from the peptic fragments during peptic digestion and HPLC is unavoidable
in these experiments, simple adjustments can be made for these
losses.
These adjustments are based on analysis of appropriate
controls (Zhang, 1995; Zhang & Smith, 1993): |
| a
0%
deuterium control |
|
and |
a 100% deuterium
control
|
|
|
Control samples representing 0% and 100%
exchanged protein are analyzed with each set of samples. The 0% control is
protein that had never been exposed to deuterium and the 100% control is
protein in which all exchangeable hydrogens have been replaced with
deuterium. The adjustment for loss of deuterium is explained with
the help of Figure 6. |
| |
 |
 |
 |
|
Figure
6 --
Example of data processing steps for adjusting the value of
incorporated deuterium. (a). The centroid value of each isotopic distribution representing a
given peptide is determined and adjusted with the equation in
(b) for loss of deuterium during analysis. See text for
definition of terms in (b). Panel (c) compares raw data
for a segment of protein. Observed: measured amount of
deuterium taken-up at the given time; Adjusted: amount of
deuterium in the peptide after adjusting the data with the
equation in (b); Calculated: amount of deuterium in an
unstructured peptide with the sequence of this segment of
protein under identical exchange-in
conditions. |
|
| |
|
Panel a is an example of data for a
segment of protein, displaying the 0% and 100% references. The equation
used to apply the adjustment is shown in Figure 6b where D is the adjusted
deuterium level, m is the experimentally observed mass, m0% is the 0% or undeuterated control, m100% is the totally deuterated control and N is the
total number of amide hydrogens in the fragment. |
| |
|
Since the adjustment is based on the
assumption that loss of deuterium from a totally deuterated fragment is
proportional to the loss from the same fragment when partially deuterated,
the accuracy of the adjustment depends on the sequence of the peptide
(Zhang & Smith, 1993). The average error that results from the
equation in Fig 6a is about 5%, with 92% of fragments having an error
<10%, 3% having an error >15% and 0.5% having an error >25%
(Zhang & Smith, 1993). |
| |
|
Figure 6c shows the deuterium level
plotted versus exchange-in time for a segment of protein before and after
the adjustment for back-exchange was considered. The absolute number of
deuterium is affected, but the profile of the curve is not. For
comparison, a curve for exchange of the same sequence in an unstructured
peptide [calculated from the parameters in (Bai et al., 1993)]. |
| |
|
Adjustment is most useful when
determining the actual amount of deuterium incorporated in a fragment.
However, when the levels of deuterium in the same fragment are compared
(such as comparing the amount of deuterium in a given region in the free
or ligand bound state) the correction is not required as long as the
analysis conditions are identical. |
| |
| |
| Calculation
of rate constants |
| |
|
From exchange-in
time-course data, one can estimate the distribution of rate
constants describing isotope exchange at each linkage within a
peptide. An example of fitting the data is shown in
Figure 7. |
| |
|
|
| |
|
Real data for a fragment of a protein are
shown in panel a. A smooth line corresponding to the equation that best
fits the experimental data was calculated using the equation in panel b.
Rate constants are fit to the experimental data with a series of first
order rate expressions where D is the number of deuterium present in a
peptide, N is the number of peptide amide linkages in a segment and ki are the hydrogen-deuterium exchange rate
constants for each peptide linkage (Zhang et al., 1996). |
| |
|
Although single amino acid resolution is
not generally possible, exchange rates describing groups of hydrogens with
similar exchange rates can be calculated. Therefore, a three-term
exponential equation was used where the amide hydrogens are divided into
fast, medium and slow categories according to their exchange
rates |
| |
|
Figure 7c shows the actual equation used
for the data in Fig 7a. The number of amide hydrogens that exchange at
similar exchange rates is calculated for each term of the equation. At the
same time the average rate constant for each group is calculated. The
results for fitting the experimental data are shown in the bottom of panel
c where the number of amide hydrogens that exchange at a given rate are
tabulated along with the average rate constant calculated for each
group |
| |
|
Software packages such as Kaleidograph
(Synergy Software) or SigmaPlot (Jandel) can be used for
fitting. |
| |
|
|
References |
| |
|
Bai, Y., Milne, J. S., Mayne, L., &
Englander, S. W. (1993). Primary structure effects on peptide group
hydrogen exchange. Proteins: Struct. Funct. Genet.
17, 75-86. |
| |
|
Zhang, Z., Post, C. B., & Smith, D.
L. (1996). Amide hydrogen exchange determined by mass spectrometry:
application to rabbit muscle aldolase. Biochemistry
35, 779-791. |
| |
|
Zhang, Z., & Smith, D. L. (1993).
Determination of amide hydrogen exchange by mass spectrometry: A new tool
for protein structure elucidation. Protein Sci.
2, 522-531. |